Main Article Content
Evaluation of design parameters of eight dental implant designs: A two‑dimensional finite element analysis
Abstract
Materials and Methods: A two‑dimensional finite element model of an implant–bone system is developed by using Ansys. An oblique load of 100 N 45° to the vertical axis of implant as well as a vertical load was considered in the analyses. The study evaluated eight types of different thread designs to evaluate stresses and strains around the implants placed in D1 bone quality.
Results: Forty‑five‑degree oblique von Mises stresses and strains were the highest for the filleted and rounded square thread with an angulation of 30° (216.70 MPa and 0.0165, respectively) and the lowest for the trapezoidal thread (144.39 MPa and 0.0015, respectively).
Conclusions: The findings in this study suggest that the filleted and rounded square thread with an angulation of 30° showed highest stresses and strains at the implant–bone interface. The trapezoidal thread transmitted least amount of stresses and strains to the cortical bone than did other models.