Main Article Content

Determination of Total Iron as Fe(II) in Multivitamins, Haematinics and Natural Waters using a Sequential Injection (SIA) System


JF Van Staden
EB Naidoo

Abstract

The determination of total iron in pharmaceutical products and natural waters as Fe(II) using a sequential injection system was investigated. A cadmium reductor consisting of cadmium granules was used to reduce Fe (III) to Fe (II). The Fe (II) was then determined (by its reaction with 1,10 Phenanthroline) as a [Fe(phen)32+] complex at 515 nm with a UV/Vis spectrophotometer. The linear range of the system is between 1 and 60 mg/l with a detection limit of 0.18 mg/l. The proposed system is suitable for the determination of total iron as Fe(II) in pharmaceutical products and natural waters at a rate of 24 samples/hour with a relative standard deviation of less than 2.5%. Statistical comparison between the proposed sequential injection (SIA) system, certified values and the standard methods (Inductively Coupled Plasma {ICP} and UV/Vis spectrophotometry) revealed that there is no significant difference at the 95% confidence level.


South African Journal of Chemistry Vol.53(3) 2000: 191-205

Journal Identifiers


eISSN: 1996-840X
print ISSN: 0379-4350